a-AB23PH Quick Start Guide
Switch On/O󰀨
On
O󰀨
Switch pH/mV Mode
Main screen pH mode mV mode
Switch Endpoint Mode
Auto end point mode Continuous reading mode
Bu󰀨er Group Setup
Temperature Setup
Note: Predened buffer group 1) 1.68 4.01 7.00 10.01 12.45 2) 1.68 4.01 6.86 9.18 12.45 (25°C)
Select the bu󰀨er group
Conrm
Conrm Conrm
Calibration
Measurement
Store/Recall Data
Calibrating (1
st
point) Completed Completed (2
nd
point) completed
Note: after the 2
nd
calibration point is
completed, press the Store button to end
the calibration or follow the previous steps to
perform a 3
rd
point calibration.
Insert a probe into the 1st
bu󰀨er solution.
Insert the probe into a sample.
Measuring Completed
Store measurement data:
After the measurement is
completed, press the Store
button.
Recall measurement data:
press and hold the Store
button to recall.
Recall calibration data:
After the calibration is
completed, press and hold
the Cal button to recall.
(
°C
/
°F)
Setup temperature unit
Setup MTC temperature