S
Shannon BauerAug 12, 2025
What does 'CAL E' mean on OHAUS Defender T24PE Measuring Instruments?
- JJuan EllisAug 12, 2025
The 'CAL E' message on OHAUS Measuring Instruments indicates that the calibration value is outside of allowable limits.
What does 'CAL E' mean on OHAUS Defender T24PE Measuring Instruments?
The 'CAL E' message on OHAUS Measuring Instruments indicates that the calibration value is outside of allowable limits.
Why is the battery icon flashing on my OHAUS Defender T24PE?
The battery icon flashes on your OHAUS Measuring Instruments because the battery is empty.
What does Error 8.4 mean on OHAUS Defender T24PE Measuring Instruments?
Error 8.4 on OHAUS Measuring Instruments means the weight reading is below the underload limit.
What does 'rEF.Err' mean on OHAUS Defender T24PE Measuring Instruments?
The 'rEF.Err' message on OHAUS Measuring Instruments indicates that the reference weight is too small. Specifically, the weight on the pan is too small to define a valid reference weight.
What does Error 8.1 mean on OHAUS Defender T24PE Measuring Instruments?
Error 8.1 on your OHAUS Measuring Instruments indicates that the weight reading exceeds the Power On Zero limit.
What does Error 8.2 mean on OHAUS Defender T24PE?
Error 8.2 on your OHAUS Measuring Instruments means that the weight reading is below the Power On Zero limit.
What does Error 8.5 mean on my OHAUS Measuring Instruments?
Error 8.5 on your OHAUS Measuring Instruments means that the device was tared at one unit, but after switching to another unit, the tare value exceeds the maximum.
What does 'Lo.rEF' mean on OHAUS Measuring Instruments?
The 'Lo.rEF' message on OHAUS Measuring Instruments is a warning that the Average Piece Weight is too small.
Capacity | 24 kg |
---|---|
Display | LCD |
Housing Material | ABS plastic |
Protection | IP54 |
Calibration | External calibration |
Interface | RS232 |
Power Supply | AC adapter |
Operating Temperature | 0°C to 40°C |
Units of Measurement | g, kg, lb, oz |
Approvals | CE |
Functions | Weighing |
Outlines critical safety guidelines for the safe installation and operation of the indicator.
Explains automatic and manual zeroing procedures for accurate measurements.
Details how to use the tare function to exclude container weight.
Lists error codes, their descriptions, and potential causes for troubleshooting.
Details the technical specifications of the T24PE indicator, including capacity, resolution, and power.