What does Error 9.5 mean on OHAUS Measuring Instruments?
- Jjennifer45Sep 9, 2025
If your OHAUS Measuring Instruments displays Error 9.5, it indicates that calibration data is not present. Calibrate the scale.
What does Error 9.5 mean on OHAUS Measuring Instruments?
If your OHAUS Measuring Instruments displays Error 9.5, it indicates that calibration data is not present. Calibrate the scale.
How to fix OHAUS Scales that won't zero?
If you cannot zero your OHAUS Scales, first, remove any weights from the pan that exceed the allowable limits. Ensure the weight is stable by waiting for any vibration or movement to cease, or increase the filtering. If the issue persists, service may be required, as the load cell may be damaged.
What does Error 8.1 mean on OHAUS Scales?
If your OHAUS Scales display Error 8.1, it indicates that the weight reading exceeds the Power On Zero limit. Remove weight from the pan. If the issue persists, recalibrate the scale.
What does error 8.4 mean on OHAUS Scales?
If your OHAUS Scales show Error 8.4, the weight reading is below the Underload limit. Add weight to the pan. If the problem continues, recalibrate the scale.
What does Error 8.93 mean on OHAUS Measuring Instruments?
If your OHAUS Measuring Instruments displays Error 8.93, the EasyConnectTM base is connected to a blank indicator under the LFT ON status. Break the seal, connect the base to the indicator, and power on the scale. Power off the scale and perform sealing again.
What does Error 8.1 mean on my OHAUS Accessories scale and how do I fix it?
Error 8.1 on your OHAUS Accessories scale indicates that the weight reading exceeds the Power On Zero limit. Remove weight from the pan. If the problem persists, recalibrate the scale.
What does Error 8.3 mean on OHAUS Measuring Instruments?
If your OHAUS Measuring Instruments displays Error 8.3, it means the weight reading exceeds the Overload limit. Reduce the weight on the pan.
What does Error 8.1 mean on OHAUS Measuring Instruments?
If your OHAUS Measuring Instruments displays Error 8.1, it indicates the weight reading exceeds the Power On Zero limit. Remove weight from the pan. You may also need to recalibrate the scale.
Why is my OHAUS Measuring Instruments unable to calibrate?
If you are unable to calibrate your OHAUS Measuring Instruments, ensure that the LFT security switch is turned off. Also, verify that you are using the correct calibration mass.
What does CAL E mean on OHAUS Measuring Instruments?
If your OHAUS Measuring Instruments displays CAL E, the calibration value is outside allowable limits. Use the correct calibration weight.
Capacity | 33 kg |
---|---|
Readability | 0.1 g |
Display | LCD with backlight |
Power Supply | AC adapter or rechargeable battery |
Approval | OIML, NTEP |
Application | Weighing, Parts Counting, Checkweighing |
Construction | Stainless Steel |
Calibration | External calibration |
Communication | RS232, USB |
Design Features | Overload protection |
Units | g, kg, lb, oz |
Operating Temperature | -10°C – 40°C |