E
emillerSep 9, 2025
What does Error 9.5 mean on OHAUS Measuring Instruments?
- Jjennifer45Sep 9, 2025
If your OHAUS Measuring Instruments displays Error 9.5, it indicates that calibration data is not present. Calibrate the scale.
What does Error 9.5 mean on OHAUS Measuring Instruments?
If your OHAUS Measuring Instruments displays Error 9.5, it indicates that calibration data is not present. Calibrate the scale.
Why is my OHAUS Measuring Instruments unable to calibrate?
If you are unable to calibrate your OHAUS Measuring Instruments, ensure that the LFT security switch is turned off. Also, verify that you are using the correct calibration mass.
Why is the battery symbol flashing on my OHAUS Measuring Instruments?
If the battery symbol is flashing on your OHAUS Measuring Instruments, it indicates a low battery. Replace the dry batteries for the i-DT33P model or recharge the battery for the i-DT33XW model.
What does Error 8.92 and Error 8.94 mean on OHAUS Measuring Instruments?
If your OHAUS Measuring Instruments displays Error 8.92 or Error 8.94, it indicates incorrect data from the EasyConnectTM base under LFT ON status. Break the seal or replace the original base/indicator. Service may be required.
What does Error 8.1 mean on my OHAUS Accessories scale and how do I fix it?
Error 8.1 on your OHAUS Accessories scale indicates that the weight reading exceeds the Power On Zero limit. Remove weight from the pan. If the problem persists, recalibrate the scale.
What does error 8.8 mean on OHAUS Scales?
If your OHAUS Scales display Error 8.8, the factory calibration data in the EasyConnectTM module on the load cell cable is not valid under LFT off status. Calibrate the scale to resolve this.
What does Error 8.2 mean on OHAUS Scales?
If your OHAUS Scales display Error 8.2, it means the weight reading is below the Power On Zero limit. To fix this, add weight to the pan. If the issue remains, recalibrate the scale.
What does error 8.4 mean on OHAUS Scales?
If your OHAUS Scales show Error 8.4, the weight reading is below the Underload limit. Add weight to the pan. If the problem continues, recalibrate the scale.
What does Error 8.1 mean on OHAUS Scales?
If your OHAUS Scales display Error 8.1, it indicates that the weight reading exceeds the Power On Zero limit. Remove weight from the pan. If the issue persists, recalibrate the scale.
How to fix OHAUS Scales that won't zero?
If you cannot zero your OHAUS Scales, first, remove any weights from the pan that exceed the allowable limits. Ensure the weight is stable by waiting for any vibration or movement to cease, or increase the filtering. If the issue persists, service may be required, as the load cell may be damaged.
Capacity | 33 kg |
---|---|
Readability | 0.1 g |
Display | LCD with backlight |
Power Supply | AC adapter or rechargeable battery |
Approval | OIML, NTEP |
Application | Weighing, Parts Counting, Checkweighing |
Construction | Stainless Steel |
Calibration | External calibration |
Communication | RS232, USB |
Design Features | Overload protection |
Units | g, kg, lb, oz |
Operating Temperature | -10°C – 40°C |