35
BX61WI
3 Filter Characteristics of Fluorescence Mirror Units
Excitation
Method
U-MGFP/XL
BA510IF
IB
U-MGFPA/XL
DM505 BP460-490
BA510-550
For EGFP, S65T, RSGFP.
(U-MGFPA/XL is for fluorochrome
separation.)
Mirror Unit Dichroic Mirror Excitation Filter Barrier Filter Application
1. IB excitation (wide bandwidth) U-MGFP/XL 2. IB excitation (wide bandwidth) U-MGFPA/XL
Transmittance (%)
Transmittance (%)
Wavelength (nm) Wavelength (nm)
4 Fabricating Optional Mirror Units
}An optional mirror unit can be fabricated by attaching the custom-order barrier filter, excitation filter and dichroic mirror to
the U-MF/XL.
Dimension Conditions of Optical Components of Mirror Unit
· Barrier filter: 32
-0.1/-0.2
mm, max. thickness 4 mm
· Excitation filter: 25
-0.1/-0.2
mm, max. thickness 6 mm
· Dichroic mirror: See figure on the right.
#When replacing the dichroic mirror, take special care not to stain it by leaving fingerprints, etc.
UP
Dichroic mirror
(made to custom order)
Interference mirror surface
Thickness: 1.5 ±0.05 mm
Barrier filter (made to custom order)
Excitation filter
(made to custom order or
ommercially marketed product)