26
Item
Specifications
CK30 CK40
8. Stage Stage dimensions: 160 (W) X 250 (D) mm
Stage extension plate dimensions: 70 (W) X 180 (D) mm
Provided with ø35 mm petri dish holder (CK30 only)
9. Mechanical stage
Traversing area: 120 (X) X 78 (Y) mm; coaxial low drive control knobs on right or
left side of plain stage; provided with 3 culture vessel holder
10. Condenser
Ultra-long-working distance condenser N.A. 0.3, W.D. 72 mm; detachable
11. Dimensions and weight
236 (W) X 469 (D) X 476 (H) mm
8 kg (17.6 lb)
12. Operating environment
· Indoor use
· Altitude : Max. 2,000 m
· Ambient temperature: 5°C to 40°C (41°F to 104°F)
· Maximum relative humidity: 80% for temperatures up to 31°C (88°F) decreas-
ing linearly through 70% at 34°C (93°F), 60% at 37°C (99°F), to 50% relative
humidity at 40°C (104°F)
· Main supply voltage fluctuations not to exceed ±10% of nominal voltage
· Pollution Degree: 2 (in accordance with IEC664)
· Installation/Overvoltage Category: II (in accordance with IEC664)
236 (W) X 371 (D) X 476 (H) mm
8.6 kg (18.9 lb)