28
Problem Cause Remedy Page
The image is too dark. Exposure compensation is set to the -
side beyond the appropriate value.
Set the exposure compensation to
the appropriate value.
-
The bright area of the image is set as the
metering area erroneously.
Set the area where you want to meter
as the metering area.
-
The AE lock is set at the shorter
exposure time than the one required
currently.
Cancel the AE lock.
-
The illumination of the microscope is too
dark
Adjust the brightness by increasing
the illumination intensity of the
microscope or removing the
attenuation filter from the light path.
-
The colors of the image are
strange.
The reference area for white balance is
not selected appropriately.
Select the area where nothing is
shown on the background as the
reference area for white balance.
-
The RGB balance for manual white
balance is not adjusted appropriately.
Adjust the RGB balance appropriately.
-
The screen color of the PC is not set
appropriately.
Set the screen color of the PC to 24
bit or higher. 32 bit is recommended.
The image color is not selected
appropriately.
Select the color mode which
is suitable for your microscope,
observation method and specimens.
-