30
Problem Cause Remedy Page
The acquired image is not
in focus.
The specimen is not brought into focus. Bring the specimen into focus
accurately.
-
The parfocality between the camera
adapter and the eyepieces is not
adjusted properly.
Adjust the parfocality between the
camera adapter and the eyepieces
properly.
-
The aperture iris diaphragm of the
condenser is open too wide.
Narrow down the aperture iris
diaphragm appropriately and bring
the specimen into focus accurately.
-
The field iris diaphragm is open too wide. Narrow down the field iris diaphragm
appropriately and bring the specimen
into focus accurately.
-
Lens components of the microscope
and the cover glass on the bottom of the
camera head are dirty.
Clean the objective, condenser,
window lens of the microscope, and
the cover glass on the bottom of the
camera head.
15
The microscope and camera are
subjected to vibration during acquisition.
Acquire images in an environment
where the microscope and camera
are not vibrated. It is effective to use
an anti-vibration table.
-