DP73
47
Problem Cause Remedy Page
Still images cannot be recorded.
The camera and PC are connected
improperly.
Connect the camera head and PC
properly.
26, 31
The cellSens or OLYMPUS Stream
is processing recording.
Wait until the recording processing
completes before starting record-
ing of the next image. In certain
cases, it may be required to press
the Cancel button on the status bar
and record the image again.
(Online
manual)
The cellSens / OLYMPUS Stream is
processing file save, etc.
Wait until the processing com-
pletes before starting recording of
the next image.
(Online
manual)
The PC memory is insufficient. Exit from other software before retry-
ing recording.
–