Problem Cause Remedy Page
Picture is too dark.
Exposure correction is set in the –
direction.
Return the exposure correction
value to 0 and set the desired
exposure correction value.
(Online
manual)
The metering area is set to a bright
area outside the region of interest.
Move the metering area to the area
where you want to obtain optimum
exposure.
(Online
manual)
AE lock, which was set when the
exposure time was shorter than the
currently required exposure time, is
active.
Cancel AE lock.
(Online
manual)
The output highlight level adjust-
ment is too low.
Reset the current level adjustment
and adjust the optimum level
again.
(Online
manual)
The microscope illumination is too
dark.
Increase the microscope illumina-
tion intensity or disengage the
existing ND filter to increase bright-
ness.
–