Chapter 3 Preparation and Inspection
23
LF-TP/DP/GP OPERATION MANUAL
1. Inspect the sterile package for tears, inadequate sealing or water damage. If
the sterile package show any irregularities, the sterile condition of the
instrument may have been compromised. Use a spare instead.
2. Inspect the Biopsy Valve for cracks or other damage.
Inspection of the Biopsy Valve
Inspect the Biopsy Valve (MD-495) for damage (see Figure 3.5).
Figure 3.5
• Biopsy Valve (MD-495) is included with endoscopes shipped
to the UK and all other countries where the Single-use Biopsy
Valve (MAJ-210) is not included.
• If using a valve which is provided non-sterile, sterilize it
according to the Recommended ETO Gas Exposure
Parameters given in the “OLYMPUS ENF/LF ENDOSCOPE
REPROCESSING MANUAL”.