33
1EAS 1EAS?
Functions
1. Sets the aperture iris diaphragm to the specified position.
2. Checks the current position of the aperture iris diaphragm.
Special notes
1. The aperture iris diaphragm positions are referred to as; 0 = minimum; 1550 = intermediate; 3113 = maximum.
2. When a position out of the specified range (0 - 3113) is specified, an illegal parameter error is identified and no operation
is executed.
3. The aperture iris diaphragm controlled with these commands is the reflected light aperture iris diaphragm.
Command descriptions
Aperture Iris Switching
1EASB n
(n: 0 to 3113)
Command Type Description Response Result Description
Request Setting of the aper-
ture iris diaphragm
to the position with
the number speci-
fied with “n”.
1EAS + Successful The aperture iris diaphragm is set
to the specified position.
1EAS !,E01130
Error The connection is disconnected
due to an abnormality with the
aperture iris diaphragm.
1EAS !,E01120 Error A string other than “?” is sent af-
ter “1EAS”.
1EAS?
Query Check of the cur-
rent position of the
aperture iris dia-
phragm.
1EAS n
(n: 1 to 3113)
Successful
The current position of the aper-
ture iris diaphragm is returned.
1EAS !,E01150 Error A string is sent after “1EAS? ”.
*
)
“ ” represents a space.
Line feed command <CR><LF> is required after commands and line feed command <CR><LF> are added after any
response.