Chapter 3 Installation and Connection
VIDEO SYSTEM OTV-SI
When connecting the video adapter and endoscope, refer to
their respective instruction manuals.
Connecting the light guide plug of a videoscope or light guide
cable
• Only use compatible endoscopes. Using incompatible
endoscopes can result in patient injury and/or equipment
damage. Refer to the “System chart” in the Appendix for a list
of endoscopes that are compatible with this instrument.
• If the surface of the endoscope is dirty, reprocess it according
to the directions given in the endoscope’s operation or
reprocessing manual before connecting it to this instrument.
Otherwise, it may cause patient injury, equipment damage,
and/or improper illumination.
Confirm that the endoscope is thoroughly dry before
connecting. If the endoscope is wet (after disinfection, for
example), dry it according to the directions given in the
endoscope’s operation or reprocessing manual before
connecting it to this instrument. Otherwise, equipment
damage and/or improper illumination can result.