7.4 Ambient Conditions
7.4.1 Climate
Climate
Temperature
Operation
Storage and transport
32 ... 122°F (0 ... +50°C)
-13 ... 158°F (-25 ... +70°C)
Relative humidity
5 … 95% relative humidity; noncondensing
Table 11 Climate
7.4.2 Shock and Vibration
Dynamics
Vibration
Tested according to IEC68-2-6 (operating
mode) frequency range 10..150Hz;
acceleration 2g continuous (20m/s
2
);
10 sweeps, each axis
Shock
Tested according to IEC68-2-27
(operating mode) 15g/11ms, half-sine,
each axis
Table 12 Shock and Vibration
7.4.3 Electromagnetic Compatibility (EMC)
EMC
CE Conformity
The product complies to the normative
document about electromagnetic compatibility
for standardization of the laws of the member
states of the council of the European Union
(EMC standard 89/336/EWG).
Emission
International
Europe
FCC Subpart B of Part 15 Class A
EN 55011/3.1991/Group1/Class A
Susceptibility
International
Europe
IEC 801-2/3/4
EN 50082-2:1992
Table 13 Electromagnetic compatibility