11
SECTION 3 — BEAD RUPTOR SAMPLE TUBES
3.1 TUBES & TUBE CARRIAGES (cont.)
12 x 7 mL screw cap tubes
7 mL Tube Volume and Tissue Size:
Minimum: 1 mL, 1 g
Maximum: 5 mL, up to 4 g
CAUTION: The maximum recommended speed for the 7 mL
Tube Carriage is 6 m/s.
3 x 15 mL screw cap tubes
15 mL Tube Volume and Tissue Size:
Minimum: 2 mL, 300 mg
Maximum: 10 mL, up to 10g
CAUTION: The maximum recommended speed for the 15 mL
Tube Carriage is 7 m/s. The 15ml tube carriage is not compatible with
the BR-Cryo lid.
6 x 30 mL screw cap tubes
30 mL Tube Volume and Tissue Size:
Minimum: 4 mL, 1 g
Maximum: 30 mL, up to 20 g
CAUTION: The maximum recommended speed for the 30 mL
Tube Carriage is 5 m/s. The 30ml tube carriage is not compatible with the BR-Cryo lid
3 x 50 mL screw cap tubes
50 mL Tube Volume and Tissue Size:
Minimum: 5 mL
Maximum: 50 mL
CAUTION: The maximum recommended speed for the 50 mL
Tube Carriage is 5 m/s. The 50ml tube carriage is not compatible with the BR-Cryo lid
CAUTION: Use only Omni sample tubes. Accuracy and operating lifetime may be aected if alternate tubes
are used.