EasyManua.ls Logo

Omni Bead Ruptor 24 - Programming the Bead Ruptor; Programming Overview and Parameters

Omni Bead Ruptor 24
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
22
SECTION 5 — PROGRAMMING
5.1 PROGRAMMING
The Bead Ruptor 24 is programmable for Speed/Processing Energy, Time, Number of Runs and Dwell (or
Pause) between cycles. The Bead Ruptor memory can store up to 99 programs.
After powering on, the LCD screen will display the program number blinking to denote this is the active’
variable. The and keys move the cursor between the dierent variables within the program while
the and buttons respectively decrease or increase the value of the active variable.
Each press of the button will incrementally change the variable’s value. Holding down the or buttons
causes the variable to advance rapidly.
Active Variable
Program: XX
S
T
C
D
System Status Display
CLOSE LID or READY
SPEED m/s
CYCLE TIME
NUMBER OF CYCLES
PAUSE DWELL
PARAMETER OPERATING RANGE
Speed / Energy From 0.8 m/s to 8 m/s in increments of 0.15 m/s
Time From 0:01 to 9:59 in increments of 1 second
Number of Runs 1 to 10
Dwell/Pause between Runs 0:00 to 9:59
Storable Programs 1 to 99
If run is stopped or paused
system status will read
“PROGRAM PAUSED”
At the end of the program
“DONE” will be displayed
followed by “REMOVE
SAMPLES”
CAUTION: Wait until the unit stops and displays “DONE—REMOVE SAMPLES” before opening the lid.
PLEASE NOTE:
• Homogenization settings cannot be changed once the run is in progress
• The Bead Ruptor 24 is designed for non-stop use and does not require a cool down period between runs.
The user can pause or stop a run by pressing the “STOP button at any time. The unit displays the Cancel
information until the run has completely stopped.
PROG >04
20°C
S=4.50 C=02
T=0:10 D=0:10
READY