38
* thinnest part
• These specifications are subject to change without notice.
• This monitor is clinically investigated according to the requirements of ISO 81060-
2:2013. In the clinical validation study, K5 was used on 85 subjects for
determination of diastolic blood pressure.
• This monitor has not been validated for use on pregnant patients.
• IP classification is degrees of protection provided by enclosures in accordance
with IEC 60529. This monitor is protected against solid foreign objects of 0.5
inches (12.5 mm) diameter and greater such as a finger, and against oblique
falling water drops which may cause issues during a normal operation.
Weight Approximately 3.2 oz. (91 g) not including batteries
Dimensions
Approximately 3.6" (w) × 2.5" (h) × 0.5" (l)*
(91 mm × 63 mm × 13 mm) (not including the wrist cuff)
Measurable wrist
circumference
5.3" to 8.5" (13.5 to 21.5 cm)
Memory Stores up to 100 readings per user
Contents
Monitor, storage case, 2 “AAA” alkaline batteries,
instruction manual, quick start guide
Protection against
electric shock
Internally powered ME equipment
Applied part Type BF (wrist cuff)
Maximum temperature of
the applied part
Lower than +118.4°F (+48°C)