6. SPECIFICATIONS
Notes:
These specifications are subject to change without notice.
In the clinical validation study, the 5th phase was used on 85 subjects for
determination of diastolic blood pressure.
This device has not been validated for use on pregnant patients.
Model
BP710NVA HEM-7121-ZVA
Display LCD digital display
Measurement range Pressure: 0 to 299 mmHg
Pulse: 40 to 180 beats / min.
Accuracy Pressure: r3 mmHg or 2% of reading
Pulse: r 5% of display reading
Inflation Fuzzy-logic controlled by electric pump
Deflation Automatic pressure release valve
Measurement method Oscillometric method
IP classification IP 20
Power source 4 “AA” batteries 1.5V
Battery life Approximately 1000 measurements (using new alkaline batteries)
Operating temperature /
humidity / air pressure
50°F to 104°F (10°C to 40°C) / 15 to 90% RH / 700 to 1060 hPa
Storage temperature /
humidity / air pressure
-4°F to 140°F (-20°C to 60°C) / 10 to 95% RH / 700 to 1060 hPa
Weight Monitor : Approximately 8 7/8 oz. (250 g) not including batteries
Arm cuff : Approximately 6 oz. (170 g)
Dimensions Monitor : Approximately 4" (w) u 3 1/8" (h) u 5 1/8" (l)
(103 mm u 80 mm u 129 mm)
Arm cuff : Approximately 5 3/4" u 23 1/2" (air tube: 29 1/2")
(145 mm u 594 mm (air tube: 750 mm))
Cuff circumference 9" to 17" (22 to 42 cm)
Memory Up to 14
Contents Monitor, arm cuff, battery set, instruction manual, quick start guide
Applied part
Protection against
electric shock
Internally powered ME equipment
= Type BF