9. GUIDANCE AND MANUFACTURER’S DECLARATION
Guidance and manufacturer’s declaration - electromagnetic immunity
OMRON BPM is intended for use in the electromagnetic environment specified below. The customer
or the user of this OMRON BPM should assure that it is used in such environment.
Immunity test
IEC 60601 test
level
Compliance level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floor should be wood, concrete,
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
Not Applicable. Not Applicable.
Surge
IEC 61000-4-5
±1 kV line(s) to
line(s)
±2 kV line(s) to
earth
Not Applicable. Not Applicable.
Voltage dips, short
interruptions and
voltage variations on
power supply inputlines
IEC 61000-4-11
<5 % U
T (>95 % dip
in U
T)
for 0.5 cycle
Not Applicable. Not Applicable.
40 % U
T (60 % dip
in U
T)
for 5 cycles
70 % U
T (30 % dip
in U
T)
for 25 cycles
<5 % U
T (>95 % dip
in U
T)
for 5 sec.
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m 3 A/m
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
Note: U
T is the A.C. mains voltage prior to application of the test level.