The quality control test confirms that the test strips and meter are
working together properly, and that you are performing the test correctly.
It is important to perform this test:
Before you first use your meter.
Before using a new box of test strips.
When you suspect that the meter or test strips are not working
properly.
When you suspect that your test results are inaccurate, or if they
are inconsistent with how you feel.
When you suspect your meter is damaged.
At least once a week.
1. Insert a test strip into the strip port, contact bars end first and facing
up, to turn on the meter and display all the display segments. If the
audio option is on, the meter will beep, signaling the meter is turned
on.
2. Check the display to confirm that all the display segments turn on
(see display illustration above).
3. Following this display check, the system will
enter the test mode. The display will show the
date and time and the strip icon with the blood
sample icon blinking. The code number will be
displayed in the center of the screen. Make
sure that the code number that appears on
the display matches the code number (CODE)
on the test strip vial (or on the foil pouch). If
not, make sure to locate and insert the code chip that came with the
box of strips. If the codes still do not match, do not perform a test.
You will need a new package of test strips to perform a test.
The blinking test strip and blood drop icon indicates that the test
strip is inserted correctly and a drop of control solution can be
added.
Note: If the test strip has been inserted incorrectly, the meter will not
turn on.
Performing a Quality Control Test
CODE
m
m
o
lL
l
mg
dL
l
MEM
CODE
5. Audio Feature
The meter comes with the meter audio feature enabled. The meter
will give one short beep when it is turned on, after sufficient sample
has been applied to the test strip and when the result is ready. The
meter will sound three short beeps to sound a warning when an
error has occurred. Please check the error number on the display to
confirm what kind of error has occurred.
Press the M button to switch between turning the meter beep "On"
and "Off". Press the S button to confirm your selection. Pressing S
at this point will end the setup mode and power off the meter.