Linearity Correction
The display will read “LIN”
Press [ZERO] or [TARE] to toggle this mode On or OFF (Linearity
correction needs to be turned off to continue to calibration mode)
There is a range of (
-
6-6) for correcting the linearity error.
Please choose 1-6 calibration when the scale reads less than what the
correct weight is
Please choose
-
1-
-
6 calibration when the scale reads more than what
the correct weight is
Or choose 0 when there is no linearity correction needed
To change the linearity correction press [ZERO] or [TARE], then press
[M+] to conrm and continue on to the next setting or [UNIT WEIGHT]
to go back to the weighing mode
Zero Collection
When the scale displays “ZERO” you will see the AD value on the third
line
The AD value should be between 10000-20000
To change the zero collection press [ZERO] or [TARE]
Press [M+] to conrm and continue on to the next setting (calibration)
or [UNIT WEIGHT] to go back to the weighing mode
WEIGHT CALIBRATION
When to calibrate
When the scale is initially installed
When the scale’s location has changed
After the scale has not been used for a long period of time
How to calibrate
In the Scale Settings Mode enter the calibration mode
The screen should read “CALI”
Conrm there is no weight on the pan before entering the weight
calibration mode
Enter the value (in kg) of the weight you will be using to calibrate the
scale (
2
/3rds of the capacity is preferred)
Then set the weight on the scale and press [M+] to conrm
Press [M+] again to nish
The display will ash “PASS” when calibration is done
Press [UNIT WEIGHT] to go back to the weighing mode