47
As shown in the images, the in total 8 available sub menus are:
System Settings
Connection Settings
User Settings
Printer Settings
LIMS Settings
Email Settings
Backup Settings
Service Settings
4.3.1 System settings
Controlled by Bath
When switched ON, this function determines that the Heads in the network are
controlled by the ColdBlock (Bath), as should be the case in a CFPP or CPPP
Analyzer setup.
When switched OFF, this function determines that the Heads in the network are
not controlled by the ColdBlock, as should be the case in a CFPP or CPPP
Automator setup.
NOTE: the CFPP/CPPP Automator is intended for use in combination with an
existing manual cooling bath.
To learn more about AirSTAR CFPP/CPPP Automator, go to chapter 5 of this
manual.
Calibration interval
Sets the time for which a calibration reminder appears after the previous
calibration.
Database size
Sets the maximum amount of test results that are stored on the iPad before the
oldest result will be automatically deleted.