2
1.0 UNPACKING AND
INSTALLATION
BEFORE UNPACKING, REFER TO SAFETY NOTES PREVIOUS PAGE
1.1 UNPACKING
• The instrument should be unpacked and inspected immediately upon receipt:
• Check the instrument visually for damage, particularly if the packaging is
damaged.
• Verify the voltage and frequency marked on the instrument matches the
local power supply.
• Report any damage, shortfall or problems with the instrument to ORBIS BV
immediately.
• Retain all packaging for future use in shipping or long-term storage of the
instrument.
Area conditions
The instrument should be installed in an area that meets the following
conditions:
• A stable room temperature of around 18-25 °C.
• The instrument should be placed on a stable bench capable of supporting
min. 28 kg.
• Adjust the leveling feet until the instrument sits level and stable.
• Allow at least 20 cm free space around the instrument for ventilation. Do not
obstruct the air outlets (at back of the instrument).