6. COMMANDS AND RESPONSES
27
DISABLE_GAIN_STABILIZATION
Stops stabilization of the gain peak while data is being acquired. The gain stabilization
adjustment is held at its current value until either gain stabilization is reenabled with the
ENABLE_GAIN_STABILIZATION command or reinitialized with the
INITIALIZE_GAIN_STABILIZATION, SET_GAIN_PEAK or SET_GAIN_WIDTH
command. See also SHOW_GAIN_STABILIZATION.
DISABLE_HV
Turns off the high-voltage output on the rear panel of the DSPEC. Note that with no external
load on the high-voltage output it takes approximately 60 seconds for the high voltage to
drop to zero volts. See the section on the bias supply for more information about the high-
voltage output. See also ENABLE_HV and SHOW_HV.
DISABLE_OVERFLOW_PRESET
Disables the overflow preset for the currently selected device (see SET_DEVICE command).
Channels that receive a count when they contain 2147483647 counts, the maximum number
of counts, will roll over to zero counts if the overflow preset is disabled. See also
ENABLE_OVERFLOW_PRESET and SHOW_OVERFLOW_PRESET.
DISABLE_PZ_AUTOMATIC
Disables the automatic pole zero mode. See also ENABLE_PZ_AUTOMATIC.
DISABLE_REMOTE
Disables the recognition of commands on the alternate communication paths. If this
command is sent to the DSPEC via the mailbox communication path, it disables command
recognition on the serial path. If this command is sent via the serial path, it disables
command recognition on the mailbox communication path. See also ENABLE_REMOTE
and SHOW_REMOTE.
DISABLE_THRESHOLD_AUTOMATIC
Disables automatic positive and negative threshold determination. See
ENABLE_THRESHOLD_AUTOMATIC, SET_THRESHOLD_NEGATIVE, and
SET_THRESHOLD_POSITIVE.
DISABLE_THRESHOLD_SAMPLE
Allows InSight Oscilloscope to free-run without a trigger. See
ENABLE_THRESHOLD_SAMPLE and SET_THRESHOLD_SAMPLE.
DISABLE_ZERO_STABILIZATION
Stops stabilization of the zero peak while data is being acquired. The zero stabilization
adjustment is held at its current value until either zero stabilization is reenabled with the