DSPEC
®
Digital Gamma-Ray Spectrometer
28
ENABLE_ZERO_STABILIZATION command or reinitialized with the
INITIALIZE_ZERO_STABILIZATION, SET_ZERO_CHANNEL or SET_ZERO_WIDTH
commands. See also SHOW_ZERO_STABILIZATION.
ENABLE_ALARM
Begins the transmission of alarm responses, $E records, when an input stops counting. A $E
response record will be transmitted only when no host commands are being processed (after
a % response from a previous command and before another command is sent). Alarm
responses are enabled for the serial line and the mailbox communication paths
independently. If the command is sent to the MCB via the mailbox, then alarms will be sent
to the mailbox. If the command is sent via the serial line, then alarms will be sent via the
serial line. Alarms can be enabled for both communication paths at the same time. See also
DISABLE_ALARM and SHOW_ALARM.
ENABLE_BLRE_AUTOMATIC
Enables the automatic determination of baseline restorer time constant. See also
DISABLE_BLRE_AUTO, SHOW_BLRE, and SET_BLRE.
ENABLE_GAIN_STABILIZATION
Enables the stabilization of the gain peak by the previously selected method, either Gauss
mode or point mode (see SET_MODE_GAUSS and SET_MODE_POINT). Spectrum
stabilization is available only on device 1. See also DISABLE_GAIN_STABILIZATION,
SHOW_GAIN_STABILIZATION, and INITIALIZE_GAIN_STABILIZATION.
ENABLE_HV
Turns on the high-voltage output on the rear panel of the DSPEC. The voltage selected on
the rear panel will be present on the high-voltage output within 60 seconds after this
command is received unless the output is overloaded or the output is shut down by the high-
voltage shutdown. See also DISABLE_HV and SHOW_HV.
Execution Errors:
%131137085<CR>
The high voltage could not be enabled due to a high voltage
shutdown condition.
ENABLE_OVERFLOW_PRESET
Enables the overflow preset for the currently selected device (see SET_DEVICE command).
Channels that receive a count when they contain 2147483647 counts, the maximum number
of counts, will stop the acquisition for that channel’s device if the overflow preset is enabled.
The channel that caused the preset to complete will contain 2147483647 counts. An alarm
response record will be sent to the host if alarms are enabled for the device whose acquisition
is stopped (see ENABLE_ALARM command). See also DISABLE_OVERFLOW_PRESET