6. COMMANDS AND RESPONSES
29
ENABLE_PZ_AUTOMATIC
Enables the automatic pole zero mode. See also DISABLE_PZ_AUTO and
SHOW_PZ_AUTO.
ENABLE_REMOTE
Enables the recognition of commands on the alternate communication path. If this command
is sent to the DSPEC via the mailbox it enables command recognition on the serial line. If
this command is sent via the serial line it enables command recognition in the mailbox.
ENABLE_THRESHOLD_AUTOMATIC
Enables automatic determination of the positive and negative thresholds. See also
DISABLE_THRESHOLD_AUTO, SHOW_THRESHOLD_AUTO,
SET_THRESHOLD_NEGATIVE, and SET_THRESHOLD_POSITIVE.
ENABLE_THRESHOLD_SAMPLE
Enables the trigger threshold in InSight mode. See also
DISABLE_THRESHOLD_SAMPLE, SET_THRESHOLD_SAMPLE, and
SHOW_THRESHOLD_SAMPLE.
ENABLE_ZERO_STABILIZATION
Enables the stabilization of the zero peak by the previously selected method, either Gauss
mode or point mode (see SET_MODE_GAUSS and SET_MODE_POINT). See also
DISABLE_ZERO_STABILIZATION, SHOW_ZERO_STABILIZATION, and
INITIALIZE_ZERO_STABILIZATION.
INITIALIZE_GAIN_STABILIZATION
Resets the gain peak stabilization adjustment to unity (no adjustment). This value is reported
as 2048 by the SHOW_GAIN_ADJUSTMENT command. See also
SET_GAIN_ADJUSTMENT, ENABLE_GAIN_STABILIZATION, and
DISABLE_GAIN_STABILIZATION.
INITIALIZE_ZERO_STABILIZATION
Resets the zero peak stabilization adjustment to unity (no adjustment). This value is reported
as 2048 by the SHOW_ZERO_ADJUSTMENT command. See also
SET_ZERO_ADJUSTMENT, ENABLE_ZERO_STABILIZATION, and
DISABLE_ZERO_STABILIZATION.
OPEN_FILE
filename.ext
A filename is needed to open a file. Begins the download of a file to the RAM Disk by
opening
filename.ext
on the RAM Disk. See WRITE_FILE and CLOSE_FILE.