OPEN_FILE_FLASH
Begins the download of a new FLASH (ROM DISK). This command removes all files on the
RAM Disk except
autoexec.bat
and
ini83905.exe
to make room for
ROMDISK.ABS
. It
also stops all acquisitions which are in progress. See WRITE_FILE_FLASH and
CLOSE_FILE_FLASH.
PAUSE_INPUT
[input-num]
Waits for the next transition on the Sample Ready input or the beginning of the next
command before responding with a % response record. If input-num is present, it must be 0.
This parameter is provided for compatibility with other ORTEC modules.
Responses:
%000000069<CR>
Transition on Sample Ready input occurred.
%130131041<CR>
Command was aborted by beginning of next command.
PAUSE_INPUT_HIGH
[input-num]
Waits for a high level to be detected on the Sample Ready input or for the beginning of the
next command before responding with a % response record. The input level must remain
high until the MCB responds; otherwise, it may not be detected. If input-num is provided, it
must be zero. This parameter is provided for compatibility with other ORTEC modules.
Responses:
%000000069<CR>
High level was detected on the input.
%130131041<CR>
Command was aborted by beginning of next command.
PAUSE_INPUT_LOW
[input-num]
Waits for a low level to be detected on the Sample Ready input or for the beginning of the
next command before responding with a % response record. The input level must remain low
until the MCB responds; otherwise, it may not be detected. If input-num is provided, it must
be zero. This parameter is provided for compatibility with other ORTEC modules.
Responses:
%000000069<CR>
Low level was detected on the input.
%130131041<CR>
Command was aborted by beginning of next command.
REBOOT
Reboots the MCB. This command is not normally used and may terminate communication.
MCB returns a 000130.
RESET
Resets the DSPEC to the state just after power is applied. This command responds with a %
response that indicates power-up just occurred.