DSPEC
®
Digital Gamma-Ray Spectrometer
36
Execution Errors:
%131135083<CR>
The command was attempted while spectrum acquisition was in
progress. No action was taken.
SET_LLD
chan
Sets the lower level discriminator to chan, which must be between 0 and 16383. See also
SHOW_LLD.
SET_LOCATION_SAMPLE
num
Sets which location in the digital filter will be displayed in InSight mode. Normally, num is
set to 3 to sample the output of the digital filter. Setting num to 02 samples intermediate
points in the filter for debugging purposes. See also SHOW_LOCATION_SAMPLE.
SET_MODE_GAUSS
Sets the method of stabilization for both gain and zero stabilization peaks to Gauss mode.
The maximum peak width is 256 channels in Gauss mode, so that either the gain or zero (or
both) peak(s) may be reduced in width. See also SET_MODE_POINT.
SET_MODE_PHA
The mode is set to PHA.
SET_MODE_POINT
Sets the method of stabilization for both gain and zero stabilization peaks to point mode. See
also SET_MODE_GAUSS.
SET_MODE_TEST
value
Sets the test mode to value. Value = 0 exits test mode, 1 is ADC test mode, and 2 is DSP test
mode.
SET_MODE_SAMPLE
Starts InSight Oscilloscope mode.
SET_NETWORK_ID
id
Establishes a new network identifier which is to be used by the MCB for ethernet
communication. The setting only takes effect after a REBOOT. id should be no more than 15
characters. See also SHOW_NETWORK_ID.
SET_NETWORK_ADDRESS
[company,]addr
Establishes the Ethernet address used by the 83905 Ethernet chip. The company portion of
the address is usually not included causing the ORTEC ID (41020) to be used. The addr
portion should not be the same on any two MCBs connected to the network. This address is