6. COMMANDS AND RESPONSES
37
assigned at the factory and should normally never be changed. See also SHOW_
NETWORK_ADDRESS.
Execution Errors:
%131128085<CR>
The number must be between 0 and 16777215.
SET_OUTPUT
port, value
This sends the value to the port. The port number can be 0 or 1; 0 is the change sample
output and 1 is the serial port. Value must be between 0 and 255. See also SHOW_OUTPUT.
SET_OUTPUT_HIGH
[output-num]
Sets the Change Sample output to a high level. If output-num is provided, it must be zero.
This parameter is provided for compatibility with other ORTEC modules.
SET_OUTPUT_LOW
[output-num]
Sets the Change Sample output to a low level. If output-num is provided, it must be zero.
This parameter is provided for compatibility with other ORTEC modules.
SET_PAGE
num
Debugging command that maps the DPM to page num; num can range from 14. See also
SHOW_PAGE.
SET_PEAK_DELAY
value
Sets the peak detect delay parameter to value. Value can range from 07; it is normally set
to 2. See also SHOW_PEAK_DELAY.
SET_PEAK_LENGTH
value
Sets the peak detect length parameter to value. Value can range from 07; it is normally set
to 3. See also SHOW_PEAK_LENGTH.
SET_PEAK_PRESET
count
Sets the ROI peak preset to the specified count. During data acquisition when the contents of
any channel of a device that has the ROI flag set reaches the peak preset count, the preset is
complete and the acquisition is stopped. The actual number of counts in the ROI peak may
exceed the preset value by a small number of counts due to the pipelined architecture of the
DSPEC. Setting a peak preset to 0 counts disables the preset. The peak preset may be set to
from 0 (disabled) to 2147483647 counts. See also CLEAR_PRESETS and
SHOW_PEAK_PRESET.
Execution Errors:
%131135083<CR>
The command was attempted while spectrum acquisition was in
progress. No action was taken.