6. COMMANDS AND RESPONSES
39
SET_SEGMENT
number
Provided for compatibility with Model 918-type MCBs. This command has no effect for the
DSPEC. The segment number may be any value from 1 through 16.
Execution Errors:
%131128085<CR>
The specified segment number was either zero or a value greater
than 16.
SET_SHAP_CUSP
value
Sets the cusp factor to value. Valid values are 0.2–1.0. See also SHOW_SHAP_CUSP.
SET_SHAP_FLAT
value
Sets the width of the flattop to value. Value is in microseconds, ranging from 0.8–2.0 in steps
of 0.4. See also SHOW_SHAP_FLAT.
SET_SHAP_RISE
value
Sets the risetime to value. Value is in microseconds, ranging from 0.8–25.6 in steps of 0.8.
See also SHOW_SHAP_RISE.
SET_SNUM
“serialnumber”
Stores the serial number of the unit in non-volatile memory inside the MCB. The serial
number must be 8 characters or less. See also SHOW_SNUM.
SET_THRESHOLD_NEGATIVE
value
Sets the negative threshold to value. The negative threshold is normally set automatically by
the DSPEC. See also ENABLE_THRESHOLD_AUTOMATIC and
SHOW_THRESHOLD_NEGATIVE.
SET_THRESHOLD_POSITIVE
value
Sets the positive threshold to value. The positive threshold is normally set automatically by
the DSPEC. See also ENABLE_THRESHOLD_AUTOMATIC and
SHOW_THRESHOLD_POSITIVE.
SET_THRESHOLD_SAMPLE
value
Sets the sample threshold to value. In InSight mode the sample threshold is used as the trigger
level. See also SHOW_THRESHOLD_SAMPLE.
SET_TIME
hour,min,sec
Sets the time stored in the battery backed-up system clock to the specified values. hour can be
any value from 0–23; min and sec can be any value from 0–59. The current date and time are
stored for a device when an acquisition is started. See also SHOW_TIME, SET_DATE,
SHOW_DATE, SET_TIME_START, and SHOW_TIME_START.