DSPEC
®
Digital Gamma-Ray Spectrometer
40
Execution Errors:
%131137085<CR>
The time could not be set due to a hardware malfunction. Hardware
service may be required.
SET_TIME_START
hour,min,sec
Sets the start time to the specified values. Normally the start date and time are set
automatically whenever a device is started with the START command. See also
SHOW_TIME_START, SET_DATE_START, SHOW_DATE_START, SET_DATE, and
SET_TIME.
SET_TRUE
ticks
Sets the true-time counter to the specified number of ticks. The number represents true time
in units of 20 milliseconds (50 ticks per second). Normally this value is set by the DSPEC
during data acquisition. See also CLEAR_COUNTERS and SHOW_TRUE.
Execution Errors:
%131135083<CR>
The command was attempted while spectrum acquisition was in
progress. No action was taken.
SET_TRUE_PRESET
ticks
Sets the true-time preset to the specified number of ticks. During data acquisition when the
true-time counter reaches the preset number of ticks, the preset is complete and the
acquisition is stopped. Setting a true-time preset to 0 ticks disables the preset. See also
CLEAR_PRESETS and SHOW_TRUE_PRESET.
Execution Errors:
%131135083<CR>
The command was attempted while spectrum acquisition was in
progress. No action was taken.
SET_UNCERTAINTY_PRESET
percent
Sets the uncertainty preset to the specified value in percent. percent is a floating point value
from 0–99.9999. See also SHOW_UNCERTAINTY_PRESET.
Execution Errors
%131128085<CR>
The value is incorrect.
%131132080<CR>
A value must be included.
SET_WIDTH
bytes
Sets the maximum number of bytes that may be sent to the host computer by the WRITE
command.