6. COMMANDS AND RESPONSES
41
Legal Commands:
SET_WIDTH 0<CR>
Width set to default (512).
SET_WIDTH 12<CR>
Width set to minimum value of 12 bytes.
... ...
SET_WIDTH 512<CR>
Width set to maximum value of 512 bytes.
SET_WINDOW
[start, length]
Sets the window of interest to the specified start channel and number of channels. The
channels of spectral data in the window of interest are affected by commands such as LEAR,
SET_DATA, and WRITE. If neither start nor length is provided, the window is set to the
maximum size allowed by the conversion gain specified. The window of interest is always set
to the maximum size after a SET_DEVICE command or a SET_SEGMENT command.
Execution Errors:
%131128085<CR>
The start channel was higher than the conversion gain.
%131129086<CR>
The length specified one or more channels that were too high
conversion gain.
%131132080<CR>
The start channel was specified without a length. If one value
is given, the other must also be given.
SET_ZERO_ADJUSTMENT
value
Sets the zero stabilization adjustment to an arbitrary value from 65535 to 65535. The total
range of the adjustment value represents ±256 channels. This adjustment is usually only
made by the gain stabilizer, and reset to 0 with the INITIALIZE_ZERO_STABILIZATION
command. See also SHOW_ZERO_ADJUSTMENT.
SET_ZERO_CHANNEL
chan
Sets the center channel for the stabilizer zero peak. If a zero channel is chosen such that the
beginning channel or ending channel would be below channel 0 or above the maximum
channel, as determined by the conversion gain, the zero peak width is reduced until the peak
fits the device boundaries. A zero channel and width must be set before zero stabilization can
be enabled. See also ENABLE_ZERO_STABILIZATION.
Execution Errors:
%131128085<CR>
The specified channel number would create a peak that was
less than the minimum width (3 channels) or would be
outside the devices range.
%131136084<CR>
The command was attempted while zero stabilization was
enabled.
SET_ZERO_PRESET
count
Sets the Gauss mode stabilization preset for the zero peak. The preset represents the minimum
number of incremental counts that must be collected in any one channel of the zero peak

Related product manuals