6. COMMANDS AND RESPONSES
49
SHOW_INTEGRAL
[start_chan,number_of_chans]
Reports the sum of the specified group of spectral data channels. If start_chan and
number_of_chans is not provided, SHOW_INTEGRAL reports the sum of all channels in the
currently selected segment that have their ROI flag set.
Responses:
$G0000000000075<CR>
Integral reported as 0.
... ...
$G4294967294131<CR>
Integral reported as 4294967294.
$G4294967295132<CR>
Integral reported as greater than or equal to 4294967295
(maximum reportable value).
SHOW_INTEGRAL_PRESET
Reports the current ROI integral preset value. For more information about the ROI integral
preset, see SET_INTEGRAL_PRESET. See also SHOW_INTEGRAL.
Responses:
$G0000000000075<CR>
Integral preset reported as 0.
... ...
$G4294967295132<CR>
Integral reported as 4294967295.
SHOW_INTEGRAL_REMAINING
Reports the current ROI integral remaining value. For more information about the ROI
integral remaining, see SET_INTEGRAL_REMAINING. See also SHOW_INTEGRAL.
Responses:
$G0000000000075<CR>
Integral remaining reported as 0.
... ...
$G4294967295132<CR>
Integral reported as 4294967295.
SHOW_LIVE
Reports the contents of the live-time counter in units of 20 milliseconds (50 ticks per second).
See also CLEAR_COUNTERS and SET_LIVE.
Responses:
$G0000000000075<CR>
Live time reported as 0 ticks.
$G0000000001076<CR>
Live time reported as 1 tick (20 milliseconds).
... ...
$G4294967295132<CR>
Live time reported as 4294967295 ticks (over 23000 days).
SHOW_LIVE_PRESET
Reports the current live-time preset in units of 20 milliseconds (50 ticks per second). See also
CLEAR_PRESETS and SET_LIVE_PRESET.