DSPEC
®
Digital Gamma-Ray Spectrometer
50
Responses:
$G0000000000075<CR>
Live-time preset reported as disabled.
$G0000000001076<CR>
Live-time preset reported as 1 tick.
... ...
$G4294967295132<CR>
Live-time preset reported as 4294967295 ticks.
SHOW_LIVE_REMAINING
Reports the current live-time remaining in units of 20 milliseconds (50 ticks per second). See
also SET_LIVE_REMAINING.
Responses:
$G0000000000075<CR>
Live-time remaining reported as disabled.
$G0000000001076<CR>
Live-time remaining reported as 1 tick.
... ...
$G4294967295132<CR>
Live-time remaining reported as 4294967295 ticks.
SHOW_LLD
Shows the lower level discriminator setting. See also SET_LLD.
Responses:
$C00050092
The lower level discriminator is 50.
SHOW_LOCATION_SAMPLE
Shows the location sample number. See also SET_LOCATION_SAMPLE.
Responses:
$A003248
Sample number 3.
SHOW_MODE
Reports the current mode of operation (PHA or Sample). See also SET_MODE_PHA and
SET_MODE_SAMPLE.
Responses:
$FPHA<CR>
PHA mode.
$FSAM<CR>
Sample mode (InSight).
SHOW_MODE_STABILIZATION
Reports the current stabilization mode of operation. See also SET_MODE_POINT and
SET_MODE_GAUSS.
Responses:
$FPOINT<CR>
$FGAUSS<CR>