DSPEC
®
Digital Gamma-Ray Spectrometer
54
Responses:
$IT<CR>
The remote is enabled.
$IF<CR>
The remote is disabled.
SHOW_ROI
Used in conjunction with the SHOW_NEXT command, SHOW_ROI reports the first
continuous group of channels that have the ROI flag set. The response is of the form
$Dsssssnnnnnccc<CR>
where sssss represents an integer number that is the number of the first channel of the “first”
group of channels that all have their ROI bit set, and nnnnn represents an integer number that
is the number of channels in the group. The SHOW_NEXT command is used to report the
“next” group of channels that all have their ROI bit set.
Responses:
$D0100000050078<CR>
First ROI group starts at chan 1000 and is 50 channels long.
$D0215000150086<CR>
First ROI group starts at chan 2150 and is 150 channels long.
$D0000000000072<CR>
No ROI groups to report.
SHOW_ROI_SAMPLE
Displays the ROI Sample setting. See also SET_ROI_SAMPLE.
Responses:
$C00001088
SHOW_ROI_UNCERTAINTY
Reports the start channel and number of channels used in the uncertainty preset calculation.
See also SET_ROI_UNCERTAINTY.
Response:
$D0700000050ccc
Calculation is performed on channels 7000–7049.
SHOW_SEGMENT
This command is for compatibility with other ORTEC MCBs. It reports the last value that
was set with the SET_SEGMENT command. Segments have no meaning in DSPECs.
Responses:
$A001246<CR>
Segment last set to 1.
SHOW_SHAP_CUSP
Shows the Cusp Factor setting. See also SET_SHAP_CUSP.
Responses:
SHAP_CUSP 000000.99999988