6. COMMANDS AND RESPONSES
53
SHOW_PEAK_LENGTH
Shows the Peak Length setting. See also SET_PEAK_LENGTH.
Responses:
$A002247
Peak Length = 2.
SHOW_PEAK_PRESET
Reports the value of the ROI peak preset. See SET_PEAK_PRESET for information about
the ROI peak preset.
Responses:
$G0000000000075<CR>
Peak preset disabled.
$G0000000001076<CR>
Peak preset set to 1 count.
... ...
$G2147483646120<CR>
Peak preset set to 2147483646 counts.
$G2147483647121<CR>
Peak preset set to 2147483647 counts.
SHOW_PZ
Displays the PZ setting. See also SET_PZ.
Responses:
PZ_ENA 002200
Pole zero enabled and set to 2200.
PZ_DIS 001000
Pole zero disabled and set to 1000.
SHOW_PZ_AUTOMATIC
Shows the state of the automatic PZ. See also SET_PZ_AUTOMATIC.
Responses:
$IT<CR>
Pole zero enabled.
$IF<CR>
Pole zero disabled (TRP).
SHOW_PZ_DELAY
Reports the PZ Delay setting. See also SET_PZ_DELAY.
Responses:
$C00075099
SHOW_RADIX
This command is for compatibility with other ORTEC MCBs. It always reports that the
number base radix for the WRITE command is binary.
Responses:
$FBIN<CR>
Number base set to binary radix.
SHOW_REMOTE
See also SET_REMOTE.