DSPEC
®
Digital Gamma-Ray Spectrometer
56
SHOW_THRESHOLD_NEGATIVE
See also SET_THRESHOLD_NEGATIVE.
Responses:
THR_NEG 0000379791
Negative threshold is 379791.
SHOW_THRESHOLD_POSITIVE
See also SET_THRESHOLD_POSITIVE.
Responses:
THR_POS 00000126811
Positive threshold is 126811.
SHOW_THRESHOLD_SAMPLE
See also ENABLE_THRESHOLD_SAMPLE, SET_THRESHOLD_SAMPLE.
Responses:
THR_DIS 00000000003
Threshold set to 3 and disabled.
THR_ENA 00000005000
Threshold set to 5000 and enabled.
SHOW_TIME
Reports the time from the battery backed-up system clock in the form
$Nhhhmmmsssccc<CR>
where hhh represents a 3-digit integer hour (0 through 23), mmm represents a 3-digit integer
minute (0 through 59), and sss represents a 3-digit integer second (0 through 59). See also
SET_TIME, SET_DATE, and SHOW_DATE.
Example Responses:
$N010054017052<CR>
Time returned 10:54:17 (10 hours, 54 min, 17 sec).
$N020013037050<CR>
Time returned 20:13:37 (20 hours, 13 min, 37 sec).
SHOW_TIME_START
Reports the time of the last START command in the form
$Nhhhmmmsssccc<CR>
where hhh represents a 3-digit integer hour (0 through 23), mmm represents a 3-digit integer
minute (0 through 59) and sss represents a 3-digit integer second (0 through 59). See also
SET_TIME_START, SET_DATE_START, and SHOW_DATE_START.
Example Responses:
$N010054017052<CR>
Time returned 10:54:17 (10 hours, 54 min, 17 sec).
$N020013037050<CR>
Time returned 20:13:37 (20 hours, 13 min, 37 sec).
SHOW_TRUE
Reports the contents of the true-time (real-time) counter in units of 20 milliseconds (50 ticks
per second). See also CLEAR_COUNTERS and SET_TRUE.