6. COMMANDS AND RESPONSES
57
Responses:
$G0000000000075<CR>
True time reported as 0 ticks.
$G0000000001076<CR>
True time reported as 1 tick (20 milliseconds).
... ...
$G4294967295132<CR>
True time reported as 4294967295 ticks (over 23000
days).
SHOW_TRUE_PRESET
Reports the current true-time (real-time) preset in units of 20 milliseconds (50 ticks per
second). See also CLEAR_PRESETS and SET_TRUE_PRESET.
Responses:
$G0000000000075<CR>
True time preset reported as disabled.
$G0000000001076<CR>
True time preset reported as 1 tick.
... ...
$G4294967295132<CR>
True time preset reported as 4294967295 ticks.
SHOW_TRUE_REMAINING
Reports the current true time (real time) remaining in units of 20 milliseconds (50 ticks per
second). See also SET_TRUE_REMAINING.
Responses:
$G0000000000075<CR>
True time remaining reported as disabled.
$G0000000001076<CR>
True time remaining reported as 1 tick.
... ...
$G4294967295132<CR>
True time remaining reported as 4294967295 ticks.
SHOW_UNCERTAINTY
Returns the current value of the uncertainty for the peak in the uncertainty preset. See also
SET_UNCERTAINTY.
Responses:
UNCE 0000000000008.5
Uncertainty of the peak is 8.5%.
SHOW_UNCERTAINTY_PRESET
Returns the current uncertainty preset setting. See also SET_UNCERTAINTY_PRESET.
Responses:
UNCE_PRES 000000000000000
No preset.
UNCE_PRES 0000000000008.5
Preset set to 8.5%.
SHOW_VERSION
Reports the DSPEC firmware Version number in the form
Fmmmm-vvv<CR>
where mmmm is a 4-character model designator and vvv is a 3-character version designator.