EasyManua.ls Logo

Osaka F 500 - Programmable Parameters Table

Osaka F 500
21 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
OSAKA USER MANUAL F 500 / F 500-RS V.1 Page 1
The interface circuit allows the connection of up to 32 instruments
on the same line.
To maintain the line in rest conditions a 120 Ohm resistance (Rt)
must be connected to the end of the line.
If the instrument is equipped with a serial interface, the parameter to
be programmed are the following present in the parameters group
]
tS” :
"t.AS" : Address of the station. Set a different number for each
station, from 1 to 255.
Note: The baud-rate are fixed at 9600 baud.
4.15 - ACCESSORIES
The instrument is equipped with a connector that allows the
connection of some accessories described as follow.
4.15.1 - PARAMETERS CONFIGURATION BY KEY USB
The device has a connector that allows you to transfer the operating
parameters from the "KEY USB" device equipped with a 5-pin
connector. The "KEY USB" device is used for serial programming of
devices must have the same configuration parameters, or to save a
copy of the programming device and to transfer it quickly. The
device has a USB input, allowing connection to a PC, with which,
through the software configuration "Universal Conf" or "Osaka Set
Up" is possible to configure operating parameters.
For more information, see the "KEY USB" device manual.
4.15.2 - X2” REMOTE DISPLAY
The device is possible to connect a remote display device via a
cable X2 can be up to 10 m start. The X2 device is powered directly
from equipment, displays temperature measured by the probe Pr1
by a 2-digit display and a half.
Refer to the user manual regarding X2 device for more information.
5 - PROGRAMMABLE PARAMETERS TABLE
Par.
Description
Range
Def.
Note
]SP
S. - parameters relative to Set Point
1
S.LS
Minimum Set Point
-99.9 ÷ S.HS
-50.0
2
S.HS
Maximum Set Point
S.LS ÷ 999
99.9
3
SP
Set Point
S.LS ÷ S.HS
0.0
4
SPE
Eco Set Point
SP ÷ S.HS
2.0
5
SPH
“Turbo” Set Point (or
ind. Heating Set Point
mod. HC )
S.LS ÷ SP
-2.0
]In
i. -parameters relative to inputs
6
i.SE
Probes Type
Pt = PTC
nt = NTC
P1 = Pt1000
Pt / nt / P1
nt
7
i.uP
Unit of measurement
and resolution (decimal
point)
C0 = °C with 1° res.
F0 = °F with 1° res.
C1 =°C with 0,1° res.
F1 = °F with 0,1° res.
C0 / F0 / C1 /
F1
C1
8
i.Ft
Measurement filter
oF ÷ 20.0
sec
2.0
9
i.C1
Pr1 Probe Calibration
-30.0 ÷ 30.0
°C/°F
0.0
10
i.C2
Pr2 Probe Calibration
-30.0 ÷ 30.0
°C/°F
0.0
11
i.C3
Pr3 Probe Calibration
-30.0 ÷ 30.0
°C/°F
0.0
12
i.C4
Pr4 Probe Calibration
-30.0 ÷ 30.0
°C/°F
0.0
13
i.CU
Measure offset on the
display
-30.0 ÷ 30.0
°C/°F
0.0
14
i.P2
Pr2 input function:
oF = No function
EP = Evaporator (1)
Au = Aux
cd = condenser
2E = Evaporator 2
oF / EP / Au /
cd / 2E
EP
15
i.P3
Pr3 input function:
oF = No function
EP = Evaporator (1)
Au = Aux
cd = condenser
2E = Evaporator 2
dG = digital input
oF / EP / Au /
cd / 2E / dG
oF
16
i.P4
Pr3 input function:
see i.P3
oF / EP / Au /
cd / 2E / dG
oF
17
i.1F
Function and function
logic of digital input
di1:
0 = No function
1= Door open
2= Door open with fan
stop
3= Door open with fan
and compressor stop
4= External “AL” alarm
5= External “AL” alarm
with deactivation of
control outputs
6=Selection of active
Set Point (SP-SPE)
7= Switch on/ off
(Stand - by)
8= “Turbo" cycle
activation
9= Remote command
of AUX output
10= Disable recording
of HACCP alarms
-17 / -16/ -15 / -
14 / -13 / -12 / -
11 / -10 / -9 / -8
/ -7 / -6 / -5 / -4
/ -3 / -2 / -1 / 0 /
1 / 2 / 3 / 4 / 5 /
6 / 7 / 8 / 9 / 10
/ 11 / 12 / 13 /
14 / 15 /16 /17
0
S U P P LY
U S B
S U P P LY A D A P TE R
1 2 V D C A C S U P P LY
U S B
S U P P L Y
T V R Y
ca b le 10 m M A X .
KEY USB
X2