Command Response Description
Caution
This command overwrites any set offset value.
Example
Measured value = +2.100 m
reference value = +1.500 m
output = +1.500 m
(offset calculated by the OTT PLS and applied to all other
measured values = +0.600 m)
Note
If the unit is changed afterwards (aOSU<value>!), round-
ing errors of ±0.001 are possible.
䊳
Setting/reading the measuring time
aOXM<value>!
aOXM!
a<value><CR><LF>
a<value><CR><LF>
S
et measuring time
Read out measuring time
a – Sensor address
<value> – bb
b – numbers
Input/output is without leading zeroes!
Value range: 1 … 60 [s]
Factory setting = 2 [s]
Note
If measuring time is ≥ 3 seconds, <value3> and
<value4> appear aditionally in the response to an aD0!
command. With 1 second measuring time, the response is
identical to 2 seconds measuring time.
䊳 4 … 20 mA interface – setting/reading the lower limit
aOPA<value>!
aOPA!
a<value><CR><LF>
a<value><CR><LF>
Setting the lower limit
Reading out the lower limit
a – Sensor address
<value> – pbbbb.eee
p – sign (+,–)
b – numbers (before the decimal point)
e – numbers after the decimal point
Input/output is without leading zeroes!
Value range: –9999.999 … +9999.999
Factory setting = +0.000
Note
If the unit is changed afterwards (aOSU<value>!), round-
ing errors of ±0.001 are possible.
䊳 4 … 20 mA interface – setting/reading the upper limit
aOPB<value>!
aOPB!
a<value><CR><LF>
a<value><CR><LF>
Setting the upper limit
Reading out the upper limit
a – Sensor address
<value> – pbbbb.eee
p – Sign (+,–)
b – Numbers (before the decimal point)
e – Numbers after the decimal point
Input/output is without leading zeroes.
Value range: –9999.999 … +9999.999
Factory setting = +0.000
Note
If the unit is changed afterwards (aOSU<value>!), round-
ing errors of ±0.001 are possible.
18