17.2.2 Warnings/error signals
Error during use
Beep signal Vibration signal Result Required action
1x long at intervals of
approx. 5seconds (if
mute mode (silent mode)
is activated, this signal is
not generated)
Overheated hydraulic unit Reduce activity.
3x long Charge level under 25% Charge battery soon.
Remaining operating time
approx. 24hours
5x long Charge level under 10% Charge battery soon
Remaining operating time
approx. 6hours
5x long 5x long repeated every
60seconds
Medium error (see
Page93)
e.g. a sensor is not oper
ational
Walking possible with restric
tions. Please note the change
in flexion resistance.
The product must be inspec
ted immediately by an author
ised Ottobock Service Center.
10x long 10x long Battery charge level 5%
After the beep and vibra
tion signals, the product
switches to empty battery
mode and then switches
off.
Charge the battery.
30x long 1 x long, 1 x short
repeated every 3seconds
Severe error/indication
of safety mode activa
tion (see Page93)
e.g. one or more sensors
are not operational.
Attempt to reset this error by
connecting/disconnecting the
battery charger.
If the error persists, use of the
product is prohibited. The
product must be inspected by
an authorised Ottobock Ser
vice Center.
Continuous Total failure
Electronic control no
longer possible. Safety
mode active or undeter
mined valve state.
Undetermined product
behaviour.
Attempt to reset this error by
connecting/disconnecting the
battery charger.
If the error persists, use of the
product is prohibited. The
product must be inspected by
an authorised Ottobock Ser
vice Center.
Error while charging the product
LED on
power sup
ply
Status LED
on battery
charger
Error Resolution
Country-specific plug adapter not fully
engaged on power supply
Check whether the country-specific plug
adapter is fully engaged on the power
supply.
Non-functional socket Check socket with another electric appli
ance.
Defective power supply The battery charger and power supply
must be inspected by an authorised
Ottobock Service Centre.
100
Appendices
Genium 3B1-3/3B1-3=ST

Table of Contents

Other manuals for Otto Bock Genium

Related product manuals