EasyManua.ls Logo

Otto Bock Genium - Operating states;error signals; Signals for operating states; Mode switching

Otto Bock Genium
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Declaration of conformity according to the applicable European directives
Serial number
Lot number
Caution, hot surface
Protect from moisture
Please note the instructions for use
Check the product settings using the corresponding adjustment software of the
OttobockDataStation.
17.2 Operating states/error signals
The prosthesis indicates operating states and error messages through beeps and vibration signals.
17.2.1 Signals for operating states
Battery charger connected/disconnected
Beep signal Vibration signal Event
3x long Charging mode started (3sec. after connecting the bat
tery charger)
1x short 1x before beep signal Self-test completed successfully, product is operational
Mode switching
INFORMATION
When mute mode (silent mode) is activated, no beep and vibration signals are generated.
INFORMATION
If the Volume parameter is set to '0' in the Cockpit app, there are no beep signals (see Page85).
Beep signal Vibration sig
nal
Additional action performed Event
1x short 1x short Mode switching using the Cockpit
app
Mode switching is performed using the
Cockpit app.
1x long 1x long Bouncing on the forefoot followed
by unloading the prosthetic leg
Bouncing pattern recognised.
1x short 1x short Weight taken off prosthetic leg
and leg kept still for 1 second
Switching to basic mode (mode1) car
ried out.
2x short 2x short Weight taken off prosthetic leg
and leg kept still for 1 second
Switching to MyMode1 (mode2) car
ried out.
3x short 3x short Weight taken off prosthetic leg
and leg kept still for 1 second
Switching to MyMode2 (mode3) car
ried out.
99Genium 3B1-3/3B1-3=ST