EasyManua.ls Logo

Otto Bock Genium - Status signals

Otto Bock Genium
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Error message Cause Correction
A current connection to the
component was interrupted
Check the following points:
Distance from the component to the device
Charge level of the component's battery
Bluetooth of the component switched on? ()
Hold the component with the sole of the foot
facing up to make the component "visible" for
2minutes.
If multiple components were stored, was the
correct component selected?
17.2.4 Status signals
Battery charger connected
LED on
power sup
ply
Status LED
on battery
charger
Event
Power supply and battery charger operational. Battery charger not yet connected to
receiver.
Battery charger connected to receiver, good connection.
This display turns off automatically after approximately one minute to avoid bother
some light at night. The charging process is not affected.
Battery charger disconnected
Beep sig
nal
Vibration
signal
Event Resolution
1x short 1x short Self-test completed successfully. Product
is operational.
3x short 3x short Maintenance note:
e.g. service interval has been exceeded,
temporary disruption of a sensor signal
Use the Cockpit app to check the next
maintenance date for the prosthesis
(see Page88). If the date has been
reached or exceeded, the prosthesis
with tube adapter, battery charger
and power supply have to be sent to
an authorised Ottobock Service Cen
ter.
Conduct the self-test again by con
necting/disconnecting the battery
charger.
If the beep signal sounds again and
the maintenance date has not yet
been reached or exceeded, the pros
thesis should be inspected by an
authorised Ottobock Service Center.
The product can be used without
restrictions. However, vibration sig
nals may not be generated.
5x long 5x long
(every
minute)
AXON tube adapter not connected while
removing inductive charger
Connect the AXON tube adapter and
then restart the knee joint by connect
ing/disconnecting the battery charger
If the beep/vibration signal is emitted
again, the product must be inspected
by an authorised Ottobock Service
Center.
102
Appendices
Genium 3B1-3/3B1-3=ST

Table of Contents