Battery charge level
During the charging process, the current battery charge level is indicated by the LEDs lit on the side of the char
ger.
LEDs 0 1 2 3 4 5
Battery charge
level
0%-10% 11%-30% 31%-50% 51%-70% 71%-90% >90%
17.3 Directives and manufacturer’s declaration
17.3.1 Electromagnetic environment
This product is designed for operation in the following electromagnetic environments:
Operation in a professional healthcare facility (e.g. hospital, etc.)
Operation in areas of home healthcare (e.g. use at home, use outdoors)
The customer or user of the product must ensure that it is operated in such an environment.
Observe the safety notices in the section "Information on proximity to certain areas" (see Page68).
The following tables outline the test levels of the tests to be carried out. The higher value applies if there is a differ
ence between the operating environments listed in the tables below.
Table 4 - casing
Interference immunity test levelPhenomenon EMC basic standard or
test procedure
Professional healthcare
facility
Environments in areas of
home healthcare
Electrostatic discharge IEC 61000-4-2 ± 8kV contact
±2kV, ±4kV, ±8kV, ±15kV air,
High frequency electro
magnetic fields
IEC 61000-4-3 3V/m
80MHz to 2.7GHz
80%AM at 1kHz
10 V/m
80MHz to 2.7GHz
80%AM at 1kHz
High frequency electro
magnetic fields in the
immediate vicinity of wire
less communication
devices
IEC 61000-4-3 See Table 9
Magnetic fields with rated
power frequencies
IEC 61000-4-8 30A/m
50Hz or 60Hz
Table 5 - alternate current gate for the supply input
Interference immunity test levelPhenomenon EMC basic standard or
test procedure
Professional healthcare
facility
Environments in areas of
home healthcare
Electrical fast transi
ents/bursts
IEC 61000-4-4 ±2kV
100kHz repetition rate
Surges
Line against line
IEC 61000-4-5 ±0.5kV, ±1kV
Surges
Line against ground
IEC 61000-4-5 ±0.5kV, ±1kV, ±2kV
Conducted interference
induced by high frequency
fields
IEC 61000-4-6 3V/
0.15MHz to 80MHz
6V in ISM frequency
bands between 0.15MHz
and 80MHz
80%AM at 1kHz
3V/
0.15MHz to 80MHz
6V in ISM and amateur
frequency bands between
0.15MHz and 80MHz
80%AM at 1kHz
103Genium 3B1-3/3B1-3=ST