16
5. Assemble the centering aligner and magnetic clamp on the impeller on the Allegro mixing
biocontainer as shown in Figure 16.
Figure 15 : Assembly of Allegro mixing biocontainer with magnetic clamp/aligner assembly
6. Place the Allegro mixing biocontainer in the tank by aligning the magnetic clamp with the
large port on the bottom.
Figure 16: Biocontainer tank assembly with round plastic tank
7. For round plastic tanks, gently pull the bottom drain tube through the drain port opening.
For other tanks, please refer to the specific instructions delivered with the tank.
8. Before filling the biocontainer, ensure that the bottom drain tube is clamped. If the
biocontainer contains an EZ-Drain, ensure the drain is fully closed and clamp the drain in
place using the provided plastic drain clip.
9. As the biocontainer starts to fill, gently pull the bottom surface of the biocontainer to
remove any wrinkles, especially near the impeller.
DO NOT exceed recommended biocontainer capacity.
DO NOT alter the tube and/or impeller configuration.