10. Disconnecting Drive Unit from Tank
1. When mixing is complete, press and hold the STOP button until top bar of the window
filled (2-3 sec); unless the run finished and rotation stopped automatically.
2. Firmly hold the drive unit handle and raise the drive unit slightly to release the latch.
Release the latch by pulling it toward the control box.
3. Carefully lower the rear wheels of the drive unit to the floor. Roll the drive unit on its rear
wheels away from the dolly.
4. Press down on the drive unit handle until the guide bearings are free from the guide rails.
5. Pull the drive unit away from the dolly until the guide bears are free from the rails.
Carefully lower the front wheel to the ground.
6. The magnetic mixer drive unit can now be wheeled to another station for use.
7. When mixing is completed and the biocontainer is completely drained, remove the
biocontainer by carefully pulling the impeller seat and the drain tube out of their
respective ports.
8. Remove the centering aligner and return to the supplied accessories box for future use.
The interface should remain in its locked position for future use.
9. Dispose of plastic biocontainer.
Always put the protective shield back on the biocontainer impeller before disposal.