1.2 Application of the Care and Use Procedures
This guide describes the care and use of Cadence Inline Concentrator (ILC) modules with
Delta and Omega 10 kDa and 30 kDa membranes. Do not apply these procedures to Pall’s
conventional tangential flow filtration (TFF) cassettes or Cadence single-pass TFF (SPTFF)
standard modules.
The care and use procedures described in this document apply to the products listed in Table 1:
Table 1
Cadence ILC module part numbers
Cadence ILC Module Part Number
Delta Membrane Omega Membrane MWCO* (kDa) Cassette Format
Membrane Area (m
2
)
ILD010T010407 ILOS010T010407 10 T01 0.065
ILD010T020407 ILOS010T020407 T02 0.13
ILD010T120407 ILOS010T120407 T12 0.7
ILD010T060407 ILOS010T060407 T06 3.5
ILD030T010407 ILOS030T010407 30 T01 0.065
ILD030T020407 ILOS030T020407 T02 0.13
ILD030T120407 ILOS030T120407 T12 0.7
ILD030T060407 ILOS030T060407 T06 3.5
* Molecular weight cut-off
1.3 Using the Cadence Inline Concentrator Module Safely
Read all safety warnings and operating instructions before using the Cadence ILC module.
Keep this care and use guide in an accessible location and make it available to users of the
Cadence ILC modules. General safety precautions include:
• Wear the appropriate personal protective equipment (PPE) and clothing including safety
glasses and gloves when working with Cadence ILC modules, systems, and chemical
solutions.
• The solutions recommended for sanitization may be hazardous or corrosive. Refer to the
Material Safety Data Sheets (available from the chemical suppliers) to learn how to use
the recommended solutions safely.
• Do not exceed the operating limits of the Cadence ILC module or process components.
• Provide sufficient space around the Cadence ILC module to work ergonomically and safely
when assembling and using the module.
• Disconnecting or dismantling Cadence ILC module components without first isolating and
depressurizing the system or setup can lead to injury from the uncontrolled release of high
pressure air and solutions. Depressurize the Cadence ILC module before dismantling
components.
• Some system components may be heavy. Take proper precautions when moving or lifting
equipment to prevent personal injury. Hoists or other lifting equipment may be needed to
assemble some Cadence product equipment.
• Wipe up spills promptly to prevent injury from contact or slipping.