5. Using the Cadence ILC Module
This section describes how to complete the procedures in each of the four parts of an SPTFF process,
and is organized as follows:
Preparation of Equipment and Materials
Part 1 — Pre-Use Conditioning: Procedures A through H
Pre-use conditioning establishes the fitness of the module for use.
Part 2 — Processing: Procedures I through J
Part 3 — Post-Use Conditioning (optional): Procedures K to O
Part 4 — Post-Use Disposal
5.1 Preparation: Equipment and Materials
Equipment
Cadence ILC module
SPTFF set-up consisting of (at minimum) a feed pressure source and feed pressure sensor,
and a permeate 1 pressure sensor. Typically, end-users also have retentate and/or permeate
flowmeters (or scales) to monitor their process.
Appropriate tubing set. Pall can provide tubing sets for your application, please contact your
Pall representative for more details
Containers to hold and collect product, water, buffer, and sanitizing solutions
Tubing and connectors:
For a Cadence system with Cadence T01 or T02 modules: Female Luer type connectors
and in. internal diameter (ID), that needs to withstand 4.1 barg (60 psig)
For a Cadence system with Cadence T12 modules: MPC quick disconnect fittings,
¼ in. ID, that needs to withstand 4.1 barg (60 psig)
For a Cadence system with Cadence T06 modules: MPX quick disconnect fittings,
½ in. ID, that needs to withstand 4.1 barg (60 psig)
Personal protective equipment as required by OSHA (U.S. Occupational Safety and Health
Administration) and your facility’s safety regulations
• pH meter
• Palltronic
®
Flowstar integrity test instrument
• Thermometer
Materials
• Product solution
Buffer solution compatible with product: 15 L/m
2
De-ionized, 0.2 μm filtered water: 40 L/m
2
(for pre-conditioning steps)
Sodium hydroxide: 45 L/m
2
(for pre-conditioning sanitization)
Delta membranes: 0.1 – 0.25 N
Omega membranes: 0.1 – 0.5 N
FOR REUSE: Post-conditioning cleaning agent, 50 L/m
2
Delta membranes: 0.1 N sodium hydroxide
Omega membranes: 0.5 N sodium hydroxide + 400 ppm sodium hypochlorite
• FOR REUSE: Storage solution, 10 L/m
2
– Delta membranes: 0.1 N sodium hydroxide (< 3 months); 5% ethanol, 1% sodium
diacetate (> 3 months)
– Omega membranes: 0.1 N sodium hydroxide
HELPFUL INFORMATION:
Solution volumes are provided as a guide. Appropriate volumes
must be determined by users for their specific applications.