EasyManua.ls Logo

Panasonic MCO-170M Series - Page 93

Panasonic MCO-170M Series
104 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
92
ALARMS, SAFETY, AND SELF-DIAGNOSIS
Alarm o
safety
function
Conditions Display Buzzer
Remote
alarm
Safety
operation
Ambient
temperature
sensor error
The ambient temperature sensor is
disconnected.
“Err09: AT Sensor Open” is displayed in
the message display field.
Intermittent
tone
ON
-----
The ambient temperature sensor is
short-circuited.
“Err10: AT Sensor Short” is displayed in
the message display field.
-----
CO
2
sensor
error
The Vref or Vgas output voltage for the
CO
2
sensor is abnormal.
“Err11: CO2 Sensor Vref Abnormal” or
“Err12: CO2 Sensor Gas Abnormal” is
displayed in the message display field.
CO
2
valve is
closed.
O
2
sensor error
The measured voltage
for the O
2
sensor is abnormal.
“Err19: O2
Sensor Abnormal” is
displayed in the message display field.
N
2
/O
2
valve
is closed.
Main heater
error
Main heater burnout occurs or the main
heater SSR is short-circuited.
“Err13: Main Heater Abnormal” is
displayed in the message display field.
-----
Bottom heater
error
Bottom heater burnout occurs or the
bottom heater SSR is short-circuited.
“Err14: Humidity Heater Abnormal”
is displayed in the message display field.
-----
Door heater
error
Door heater burnout occurs or the door
heater SSR is short-circuited.
“Err15: Door Heater Abnormal” is
displayed in the message display field.
-----
Sensor box
heater error*
1
a) High limit temperature alarm is activated.
b) The sensor box heater burnout
occurs or the sensor box heater SSR is
short-circuited.
“Err16: CO2 Box Heater Abnormal” is
displayed in the message display field.
-----
Heater SSR
burnout*
1
a) High limit temperature alarm is activated.
b) Main, bottom, door, or sensor box
heater SSR burnout occurs.
“Err17: Heater SSR Open” is displayed in
the message display field.
-----
UV lamp failure
(MCO-170MUVH/MCO-170MUV, or
when an optional MCO-170UVS is
installed)
The UV lamp burns out.
“Err18: UV Lamp Abnormal” is displayed
in the message display field.
-----
New UV lamp
replacement
(MCO-170MUVH/MCO-170MUV, or
when an optional MCO-170UVS is
installed)
The accumulated ON time reaches
approx. 5,000 h.
“Warning: UV Bulb Life” is displayed in
the message display field.
----- ----- -----
Communication
error
When communication between LCD
touch panel and control substrate is
died out or unstable.
“Err56: Communication Failure” is
displayed in the message display field.
----- ----- -----
Warming-up of
gas control
After power switch is turned ON, under
warming-up before temperature is
stable and gas control is enabled.
“Status: Gas sensor initializing” is
displayed in the message display field.
----- ----- -----
*1: After a while after the high limit temperature alarm is activated, Err16 (Sensor box heater error) and
Err17 (Heater SSR burnout) are activated.

Table of Contents

Related product manuals