Loading...
10
The bed should not be used by two or more persons.
This bed is designed for use by one person. When used by two or more persons, the bed could
be damaged, resulting in injury. User's maximum weight is 195kg.
If a physician, nurse, or caregiver needs to get on the bed for a short time, confirm the following
conditions beforehand.
The total weight of the person to get on the bed must not exceed user's maximum weight.
The mattress base has been lowered all the way, and is flat.
Do not apply more than the safe working load on the bed.
The bed could be damaged, resulting in injury. The safe working load of this bed is 2250N (230kgf).
The safe working load is the maximum load with which the bed can be operated and is the total load
of the user's weight and incidental items including the mattress and options.
Do not place the bed on the inclination which angle is 6 degree or more.
The bed may tip over, resulting in an injury.
Do not spray a spray-type insecticide directly on the bed.
There is a risk that the solvents contained in insecticide may damage, discolor, or dissolve casters
or other resin parts of the bed. Check for damage caused by insecticide because damaged or dissolved
parts may cause an unexpected injury.
Do not use electrically powered beds in the presence of flammable gases.
There is a risk of explosion.
Do not repair or modify the product or its components.
This may create unintended gaps or may cause abnormal operation, resulting in injury.
Always disconnect the power plug for the bed from the power outlet when using
electronic medical equipment.
Using the bed in conjunction with electronic medical equipment (e.g. microwave or ultrashortwave
medical equipment) may cause operational failure or malfunction.
Check for safety considerations before using the bed in conjunction with other medical equipment.
Inspect the bed regularly.
The product exhibits wear and deterioration according to the frequency and conditions of use.
Inspect for loose parts, the operation of moving parts on the bed, and the presence of damage
regularly. There is a risk of unexpected injury.
Beds damaged in accidents must be inspected and repaired before being reused.
Beds damaged in earthquakes, fires, or floods or other incidents must be inspected and repaired by
Paramount bed or your distributor. Otherwise, there is a risk of electric shock or fire due to short
circuit or current leakage from electrical parts, or injury due to malfunctioning caused by deformation
of the bed.
When disconnecting the power plug, hold the power plug body to disconnect.
When holding onto only the power cable and pulling it to disconnect the plug from the outlet,
the cable could be damaged and there is a risk of electric shock or fire due to a short circuit.
SAFETY PRECAUTIONS
2